The presented work is focused on the biomechanical study of the dental disk implant. The first part of the study deals with the strain analysis of the affected bone tissue and the dental implant loaded in the coronoapical direction by force 190N. The study includes three types of implant anchorage, four degrees (stages) of osseointegration and nine degrees describing the quality of the cancellous bone. Two types of the disk implant were solved: single disk and double disk implant. Biomechanical study of the implant was focused on a stress strain analysis of the affected bone tissue. The biggest influence on the stresses in the bone tissue has primarily an implant anchorage. By the application of correlation relationships between Young modulusand the apparent density of the bone tissue (which is measurable in patients) we can the results presented in this study express depending on this variable.

The entire article can be downloaded here:

http://www.im.fme.vutbr.cz/pdf/17_1_049.pdf

 

 